Найти
Телефон
Меню

Официальный представитель завода АО «Таурас-Феникс»

в Уфе, в Казани, в Оренбурге, в Перми, в Ижевске и в Казахстане

Taures-fenix
+7 347 201 77 10 Бесплатно по России

SU6631 – фото
SU6631 – фото
SU6631 – фото
Фото
Видео

На SU6631

Для заявки пишите в WhatsApp или Telegram

# #

Характеристики

Output function normally open / closed; (configurable); analog
Materials (wetted parts) Pipe section: stainless steel (1.4404 / 316L)
Measuring range 0.03...3.9
Electrical design PNP/NPN
Process connection 1/2" NPT external thread DN15

Все характеристики

Measuring range
0.5...65 l/min0.03...3.9 m³/h8...1030 gph0.13...17.17 gpm
Process connection threaded connection 1/2" NPT external thread DN15
System gold-plated contacts
Media ultra-pure water; water; water-based media; glycol solutions; oils; Coolants
Note on media
water-based media: for media with >10 % additives, the repeatability is the only available value
low-viscosity oils with viscosity: 7...40 mm²/s (40 °C)
high-viscosity oils with viscosity: 30...68 mm²/s (40 °C)
Medium temperature
-20...100 °C-4...212 °F
Min. bursting pressure
150 bar15 MPa
Pressure rating
100 bar10 MPa
Vacuum resistance -1000
Operating voltage 18...32 DC; (to SELV/PELV)
Current consumption < 75
Protection class III
Reverse polarity protection yes
Power-on delay time 5
Measuring principle ultrasonic
Inputs counter reset
Total number of outputs 2
Output signal switching signal; pulse signal; analog signal; IO-Link; frequency signal; diagnostic signal; totalizer switching signal
Electrical design PNP/NPN
Output function normally open / closed; (configurable)
Max. voltage drop switching output DC 2
Permanent current rating of switching output DC 100
Switching frequency DC 0...10000
Analog current output 4...20
Max. load 500
Pulse output flow rate meter
Short-circuit protection yes
Type of short-circuit protection yes (non-latching)
Overload protection yes
Measuring range
0.5...65 l/min0.03...3.9 m³/h8...1030 gph0.13...17.17 gpm
Display range
-78...78 l/min-4.68...4.68 m³/h-1236...1236 gph-20.61...20.61 gpm
Resolution
0.1 l/min0.002 m³/h1 gph0.01 gpm
Set point SP
0.9...65 l/min0.052...3.9 m³/h14...1030 gph0.23...17.17 gpm
Reset point rP
0.5...64.7 l/min0.032...3.88 m³/h8...1025 gph0.14...17.08 gpm
Analog start point ASP
-65...52 l/min-3.9...3.12 m³/h-1030...824 gph-17.17...13.74 gpm
Analog end point AEP
-52...65 l/min-3.12...3.9 m³/h-824...1030 gph-13.74...17.17 gpm
Low flow cut-off LFC
0.5...3.2 l/min0.03...0.195 m³/h8...52 gph0.13...0.86 gpm
Frequency end point, FEP
13...65 l/min0.782...3.9 m³/h207...1030 gph3.44...17.17 gpm
Frequency at the end point FRP 1...10000
Pulse length 0.002...2
Pulse value 0.02...99990000 l; 0.005...26414563.515 gal
Dynamic response T05 / T09 5,7 / 86
Response time < 0.25; (dAP = 0, T09)
Damping process value dAP 0...5
Diagnostic functions direction of flow detection; signal quality
Communication interface IO-Link
Transmission type COM2 (38,4 kBaud)
IO-Link revision 1.1.3
SDCI standard IEC 61131-9: 2013-07
Profiles
BLOBBinary Large Object transfer
Common - I&DIdentification and Diagnosis
Required master port class A
Process data analog 3
Process data binary 2
Min. process cycle time 9.6
IO-Link process data (cyclical)
Functionbit length
totalizer32
Flow monitoring32
Temperature monitoring32
status4
Output 11
Output 21
Supported DeviceIDs
Type of operationDeviceID
default1755
Ambient temperature -20...60
Storage temperature -25...80
Protection IP 67
EMC
DIN 61326-1:2021
Shock resistance
DIN IEC 68-2-2720 g (11ms)
Vibration resistance
DIN IEC 68-2-620 g (10...2000Hz)
MTTF 160
UL approval
UL approval numberI034
Pressure equipment directive can be used for group 2 fluids; group 1 fluids on request
Weight 540.4
Housing rectangular
Type of mounting inlet pipe length 5xDN; outlet pipe length 1xDN
Dimensions 148 x 48 x 55.2
Material housing: stainless steel (1.4404 / 316L); Display: PFA; sealing Display: FKM; connector: PBT
Materials (wetted parts) Pipe section: stainless steel (1.4404 / 316L)
Process connection threaded connection 1/2" NPT external thread DN15
Surface characteristics Ra/Rz of the wetted parts 49.21 µin
Display
Color display 1,44", 128 x 128 pixels
Switching function2 x LED, yellow
diagnosis1 x LED, three-color
Display unit l/min; l/h; m³/h; m/s; gpm; gph; ft/s; oz/min
Items supplied
package insert
Remarks
MW = Measured value
MEW = Final value of the measuring range
pulse and totalizer signal are only available for one of the two outputs
the accuracy indications are adhered to over the entire application area
Pack quantity 1 pcs.
Connection Connector: 1 x M12; coding: A; Contacts: gold-plated

Приемущества

Accurate flow measurement of ultrapure water, water, oils and glycol solutions
A robust, component-free measuring pipe made of stainless steel offers high media resistance and permanent ingress resistance
The highly accurate inline measurement process eliminates any influence of varying wall thicknesses and pipe materials
Operating status LED signals the sensor status and allows conclusions about process quality
Convenient operation via buttons or parameter setting via IO-Link