| Number of inputs and outputs |
Number of analog outputs: 2 |
| Measuring range |
| 0...200 l/min | 0...52.84 gpm |
|
| Process connection |
threaded connection G 1 1/4 external thread flat seal |
| System |
gold-plated contacts |
| Application |
for industrial applications |
| Installation |
connection to pipe by means of an adapter |
| Media |
water; glycol solutions; Coolants; oils |
| Note on media |
| low-viscosity oils with viscosity: 7...40 mm²/s (40 °C) | | high-viscosity oils with viscosity: 30...68 mm²/s (40 °C) |
|
| Medium temperature |
-10...80 |
| Pressure rating |
|
| Operating voltage |
19...30 DC; (to SELV/PELV) |
| Current consumption |
100 |
| Min. insulation resistance |
100; (500 V DC) |
| Protection class |
III |
| Reverse polarity protection |
yes |
| Power-on delay time |
10 |
| Measuring principle |
ultrasonic |
| Number of inputs and outputs |
Number of analog outputs: 2 |
| Total number of outputs |
2 |
| Output signal |
analog signal |
| Number of analog outputs |
2 |
| Analog current output |
4...20; (scalable) |
| Max. load |
500 |
| Short-circuit protection |
yes |
| Type of short-circuit protection |
yes (non-latching) |
| Overload protection |
yes |
| Measuring range |
| 0...200 l/min | 0...52.84 gpm |
|
| Display range |
| 0...240 l/min | 0...63.42 gpm |
|
| Resolution |
|
| Analog start point ASP |
| 0...160 l/min | 0...42.28 gpm |
|
| Analog end point AEP |
| 40...200 l/min | 10.28...52.84 gpm |
|
| Max. flow rate |
220 |
| In steps of |
|
| Dynamic response T05 / T09 |
T09 = 30 (Q > 20 l/min); (water) |
| Response time |
0.25; (dAP = 0) |
| Damping process value dAP |
0...1 |
| Ambient temperature |
-10...60 |
| Storage temperature |
-25...80 |
| Protection |
IP 67 |
| EMC |
| EN 61000-4-2 ESD | 4 kV CD / 8 kV AD | | EN 61000-4-3 HF radiated | 10 V/m | | EN 61000-4-4 Burst | 2 kV | | EN 61000-4-5 Surge | 0,5 kV | | EN 61000-4-6 HF conducted | 10 V |
|
| CPA approval |
| model number | 001US | | accuracy class | 3 | | maximum allowable error | - | | Q (min) | 0,3 m³/h | | Q (t) | 0,84 m³/h | | Q (max) | 12 m³/h |
|
| Shock resistance |
| DIN IEC 68-2-27 | 20 g (11 ms) |
|
| Vibration resistance |
| DIN IEC 68-2-6 | 5 g (10...2000 Hz) |
|
| MTTF |
203 |
| Pressure equipment directive |
sound engineering practice; can be used for group 2 fluids; group 1 fluids on request |
| Weight |
1883 |
| Housing |
rectangular |
| Dimensions |
130 x 100 x 80.5 |
| Material |
housing: AlMgSi0.5 anodized; sealing: FKM; PA 6.6; cover film: PA |
| Materials (wetted parts) |
Pipe section: stainless steel (1.4404 / 316L); NBR fiber-reinforced; FKM; PPS |
| Process connection |
threaded connection G 1 1/4 external thread flat seal |
| Display |
| Display unit | 6 x LED, green (l/min, m³/h, gpm, gph, °C, °F) | | Measured values | alphanumeric display, 4-digit | | Programming | alphanumeric display, 4-digit |
|
| Display unit |
l/min; m³/h; gpm; gph; °C; °F |
| Items supplied |
|
| Accessories (optional) |
| adapter for pipe: 1 x R 1, stainless steel, E40205 | | adapter for pipe: 1 x 1" NPT, stainless steel, E40206 |
|
| Remarks |
| sealing: only with supplied Centellen seals | | MW = Measured value | | MEW = Final value of the measuring range |
|
| Pack quantity |
1 pcs. |
| Connection |
Connector: 1 x M12; coding: A; Moulded body: brass, Optalloy-plated; Contacts: gold-plated |